Request a QUOTE!

Better systems, better service, and better healthcare are at your reach.

Overview

Siemens_logo

The Siemens Somatom Emotion 16 is incredibly compact and needs only 18 square meters of floor space.   The Emotion 16 provides high quality images with minimal life-cycle cost.  In addition, you will also have access to a wide range of clinical applications.

The Emotion 16 has been designed with a 70 cm aperture, a 50 cm scan field and a patient table that can hold patients up to 440 pounds.  The Emotion 16 offers on-the-fly reconstruction of up to 16 images per second.  A powerful and cost-effective scanner, the Emotion 16 will allow your clinic to perform routine exams as well as detailed and high resolution exams.

These units are to be sold installed as turn-key projects.

  • 16-slice acquisition
  • A wide range of clinical applications
  • Fast and efficient with 0.6 second rotation time
  • Multislice UFC detector
  • 0.75 or 1.5mm slices per rotation in less than 20 seconds
  • Small diagnostic detail for acquisition
  • 105-millisecond temporal resolution for a clear depiction of blood vessels
  • Scan any anatomy with a single breath hold
  • Pathology can be seen from any perspective
  • Low patient dose

Clinical Applications: Abdominal, Adult Cardiac, Functional Imaging, Pediatrics, Pelvis, Spinal, Thoracic

CT Technology
  • syngo LungCARE CT with NEX (Nodule Enhanced Viewing)
  • syngo Colonography with PEV (Polyp Enhanced Viewing)
MEDICAL-OUTFITTERS-28Siemens-SOMATOM-Emotion-16-1019x1030
Gantry
  • Gantry Generation: 3rd
  • Slipring: Low voltage
  • Aperture: 70 cm
  • FOV: 50 cm (standard), 70 cm (optional)
  • Tilt range: ±30 degrees
  • Type of positioning lights: Laser
  • Accuracy of positioning lights: ±1 mm
  • Focus-detector distance: 940 mm
Patient Couch/ Table
  • Couch top material: Carbon fiber
  • Couch top length and width: 222 x 43 cm
  • Horizontal movement range: 153 cm
  • Horizontal movement speeds: 1-100 mm/sec
  • Accuracy/reproducibility of table positioning: ±0.25 mm (Horizontal movement)
  • Scannable horizontal range without table top extension: 153 cm
  • Scannable horizontal range with tabletop extension(s): 153 cm
  • Vertical movement range out of gantry: 45-83 cm
  • Vertical movement range in gantry: 64-83 cm
  • Minimum couch top height outside gantry: 45 cm (Vertical movement)
  • Maximum weight allowed on couch: 200 kg (440 lbs)
  • Maximum weight on couch which still achieves stated performance specifications: 200 kg
MEDICAL-OUTFITTERS-SOMATOM-emotion-1-1600x1289
2006_Siemens_Somatom_Emotion_120
  • Type/Make: Siemens Dura 422MV Siemens Straton
  • Focal spot size(s) quoted to (X-RAY TUBE): 0.5 mm x 0.8 mm
  • IEC 336/93 standard (X-RAY TUBE) 0.7 x 0.8
  • Settings at which focal spot change automatically; small changes
  • kW = kV x mA / 1000: up to 28 kW ; large up to 45 kW
  • Total filtration (inherent + beam shaping filter) at central axis (mm Al equivalent) (X-RAY TUBE) 6.3 (140kV)
  • Anode heat capacity (MHU): 5
  • Maximum anode cooling rate (kHU/min): 810
  • Method of cooling (X-RAY TUBE) Oil to air
  • Guaranteed tube life: 200,000 scan seconds
  • Detector type: Solid state (Ultra Fast Ceramic)
  • Maximum number of simultaneously acquired data sets (no. of slices): 16
  • Number of detectors per row (Maximum anode cooling rate): 736 (1472 channels)
  • Number of elements along z-axis (Maximum anode cooling rate): 24
  • Effective length of each element at isocentre: 16 x 0.6 mm, 8 x 1.2 mm
  • Total effective length of detector array at isocentre: 19.2 mm
  • Option for more slices / rotation (Maximum anode cooling rate): Yes
Siemens_Somatom_Emotion_16_2006_TAMS_104
Siemens_Somatom_Emotion_16_2006_TAMS_106
Siemens_Emotion_16_2008_129
  • X-ray generator: VAC Power Needed
  • Type: High frequency
  • Location: Rotation assembly (Composition)
  • Power rating – kV settings available (Maximum weight allowed on the couch): 50, 80, 110, 130 kW
  • mA range and step size (Weight-bearing properties) 20-345(1mA steps)
  • Maximum mA allowed for each kV (Weight bearing properties) 80 kV: 345 mA
2009_Siemens_Somatom_CT_KeLa_110
  • Monitor/ Screen Size: 19 inches
  • Number of monitors at the console (functions of each if > 1) (Image monitor) Up to 6 (shared database)
  • Image area matrix dimensions: 1024 x 1024
  • Usual range of CT number displayed (HU) (Image monitor) (-10,240 to 30,710 if metal implants)
  • The accuracy of distance measurements in x-y plane depends on pixel size (mm)
  • Weighted CTDI (CTDIw or CTDIvol) displayed on console
  • Dose Length Product (DLP) displayed on console (Dose information)
  • Geometric Efficiency displayed on the console when 70% for all collimations
  • Control methods: Mouse, keyboard
Siemens_Emotion_16_2008_151
Scroll to Top