CT Scanner Acronyms and Terminology Explained: From kVp and mAs to CTDIvol, ASiR and XR-29

Published on November 12, 2026
GE Revolution CT: Revolution Apex, angled view, photo courtesy of GE HealthCare, with the Medical Outfitters watermark

CT has more acronyms than any other imaging modality, and most of them fall into five families: the scanner and the scan technique (kVp, mAs, pitch, DAS, FOV), radiation dose (CTDIvol, DLP, SSDE, XR-29), the image and its reconstruction (HU, MPR, MIP, FBP, the vendor names for iterative reconstruction), the exams themselves (CTA, CTPA, CCTA, LDCT, HRCT) and the IT and buying vocabulary around the room (DICOM, PACS, OEM, EOL). This glossary explains each one in plain English, with what it means for the person buying, running or servicing the scanner rather than for the physicist.

Use it as a reference: each family has its own table, the dose section explains how the terms connect to one another, and the reconstruction section decodes the brand names that GE, Siemens, Philips, Canon and Fujifilm use for the same ideas. Medical Outfitters sells, installs and services CT scanners across the United States, Puerto Rico and the Caribbean, and every term here is one our engineers and sales advisers use with customers every week.

  • Technique terms describe how the X-rays are made and the gantry moves: kVp, mA, mAs, pitch, rotation time, collimation.
  • Dose terms describe how much radiation the scan delivers: CTDIvol per slice, DLP per exam, SSDE corrected for patient size.
  • Image terms describe what you see and how it was built: Hounsfield units, windows, reformats and iterative reconstruction.
  • Exam and workflow terms name the studies and the systems that move them: CTA, LDCT, DICOM, PACS.
  • Buying and service terms decide what a scanner costs to own: scan seconds, tube coverage, EOL, XR-29.
The Map

CT Acronyms at a Glance

CT acronyms in five familiesCT acronyms in five familiesWhere each term belongsTechniquekVp, mA, mAs, AECPitch, rotation time, collimationSlices, channels, DAS, SFOV and DFOVMHU, generator kW, slip ringDoseCTDIvol, DLP, SSDE, mSvDRL and ALARARDSR and dose trackingXR-29, XR-25, Smart DoseImageHU, window width and levelMPR, MIP, MinIP, VRFBP and IR: ASiR, SAFIRE, iDose,AIDR, Intelli IPMAR, kernel, SNR, CNRExamsCTA, CTPA, CCTA, CACHRCT, LDCT, CT KUBCTU, CTC, CTECT sim, PET/CT, SPECT/CTWorkflow and buyingDICOM, PACS, RIS, HL7MWL, MPPSOEM, ISO, refurbished, EOLScan seconds, tube coverage, 510(k)Each family has its own table below.
Figure 1. The map. Most CT acronyms belong to one of five families, and knowing the family is half of knowing the term.
Family 1

The Scanner and the Scan: Technique Terms

TermStands forWhat it means for you
CT, CATComputed tomography, computed axial tomographyThe same thing; “CAT scan” is the older public name
MDCT, MSCTMultidetector, multislice CTEvery scanner sold since the early 2000s; the number that follows is the slice count
Slices and channelsReconstructed slices per rotation; physical detector channelsA 64-channel detector can reconstruct 128 slices on some platforms, so compare both; see our guide to CT scanner types by slice count
DASData acquisition systemThe electronics that read the detector; a DAS upgrade is how some scanners gain slices
kV, kVpKilovolts, kilovolt peakTube voltage; sets how penetrating the beam is. Lower kVp makes iodine brighter and raises noise
mA, mAsMilliamperes, milliampere-secondsTube current and current times time; the main dose dial
AEC, ATCMAutomatic exposure control, automatic tube current modulationThe scanner varies mA with patient thickness; a required feature under XR-29
PitchTable travel per rotation divided by beam widthAbove 1 the beam leaves gaps covered by interpolation; higher pitch is faster and lower dose
Rotation timeSeconds per gantry revolution0.25 to 0.5 s on current scanners; cardiac work needs the fastest
CollimationBeam width along the patientDetector rows times row width, for example 64 by 0.625 mm equals 40 mm coverage
Helical, spiral, axialContinuous table movement versus step and shootMost scans are helical; calcium scoring and some cardiac protocols are axial
SFOV, DFOVScan field of view, display field of viewHow much of the bore is measured, and how much is reconstructed into the image
Bore, apertureGantry opening diameter70 cm is standard, 75 to 80 cm on wide-bore systems for bariatric and radiation-therapy use
MHUMega heat unitsThe tube’s anode heat capacity; a high MHU tube runs long protocols without cooling delays
Generator kWHigh-voltage generator powerSets the maximum mA at a given kVp; 60 to 120 kW on current scanners
Slip ringThe rotating electrical contactWhat lets the gantry spin continuously; a service item
Bowtie filterA shaped beam filterFlattens the dose profile across the patient; selected per body region
DECT, spectral CT, PCCTDual-energy CT; photon-counting CTScanning at two energies to separate materials, or counting individual photons; see our note on contrast in the CT with contrast vs without guide
CBCTCone beam CTA different design with a flat-panel detector; see cone beam CT explained
GE Optima CT660 64-slice CT scanner, angled view
A 64-slice scanner such as the GE Optima CT660: 64 detector channels, a 70 cm bore and a generator in the 100 kW class. Most of the technique terms above describe how it is driven.
Family 2

Dose Terms: CTDI, CTDIvol, DLP, SSDE and XR-29

From technique to dose to imageFrom technique to dose to imageHow the numbers on the console connect1kVp and mAsVoltage and current-time set the beam;AEC varies the mA with the patient'sthickness2CTDIvolThe protocol's dose per rotation in thereference phantom, shown before everyscan3DLPCTDIvol times scan length: the dose ofthe exam, recorded in the RDSR4Effective dose and SSDEDLP converted to millisieverts; CTDIvolcorrected for the patient's size5ReconstructionFBP or iterative reconstruction turnsthe raw data into HU images; IR lowersnoise so dose can fall6Display and reformatWindows, MPR, MIP and 3D present thesame data for each questionThe dose chain is what XR-29's four features control; the reconstruction step is where the vendor brand names live.
Figure 2. From technique to dose to image. The acronyms are the links in one chain.
TermStands forWhat it means for you
CTDI, CTDIvolComputed tomography dose index, volumeThe standard dose from one rotation measured in a reference phantom, in milligray; the number on the console before every scan
DLPDose length productCTDIvol times the scanned length, in milligray-centimeters; the dose of the whole exam
SSDESize-specific dose estimateCTDIvol corrected for the patient’s actual size; closer to what the patient received
Effective dose, mSvMillisievertsA whole-body risk estimate derived from DLP; a chest CT is in the low single-digit millisieverts
DRLDiagnostic reference levelA benchmark dose for a given exam, published by the ACR and others, that departments compare their protocols against
ALARAAs low as reasonably achievableThe principle behind every dose-reduction feature on the list
RDSRRadiation dose structured reportA DICOM object the scanner creates with the dose of every exam; feeds dose-tracking software; required under XR-29
XR-29, MITA Smart DoseNEMA standard XR-29-2013Four features: DICOM dose reports, dose check alerts, automatic exposure control and reference adult and pediatric protocols. Under the Protecting Access to Medicare Act of 2014, Medicare paid 5 percent less for scans on non-compliant scanners in 2016 and 15 percent less from 2017; nearly every refurbished scanner Medical Outfitters sells is XR-29 compliant, and the ones that are not are sold with that stated
XR-25NEMA dose check standardThe alert and notification values the console shows when a protocol exceeds its dose threshold; part of XR-29
Image Gently, Image WiselyPediatric and adult dose campaignsProtocol and communication resources, not regulations
ACR accreditationAmerican College of Radiology CT accreditationPhantom image quality and dose review every three years; required by many payers

The four dose terms connect in a chain. The technique, kVp and mAs with the pitch and the filter, produces a CTDIvol for the protocol. CTDIvol multiplied by the scan length gives the DLP. DLP multiplied by a conversion factor for the body region gives an effective dose in millisieverts, and SSDE adjusts CTDIvol for the fact that a small patient absorbs more of the same beam than a large one. A buyer rarely needs the numbers, but a buyer does need the features that control them, which is what the XR-29 line is about.

Family 3

Image Terms and the Names Vendors Give Iterative Reconstruction

TermStands forWhat it means for you
HUHounsfield unitThe CT number: water 0, air about minus 1,000, dense bone several hundred and up; iodine enhancement is measured in HU
WW, WLWindow width and window levelThe display range; lung, bone and soft-tissue windows show the same data differently
MPRMultiplanar reformatCoronal, sagittal and oblique images rebuilt from the axial data
MIP, MinIPMaximum and minimum intensity projectionVessel and airway views that project the brightest or darkest voxels
VR, 3DVolume renderingThe surface-shaded pictures for surgeons and patients
FBPFiltered back projectionThe classic reconstruction; fast, noisy at low dose
IRIterative reconstructionStatistical or model-based reconstruction that lowers noise so dose can fall; sold under the brand names below
ASiR, ASiR-V, TrueFidelityGE HealthCareStatistical, then hybrid, then deep-learning reconstruction across the LightSpeed, Optima and Revolution generations
SAFIRE, ADMIRESiemens HealthineersSinogram-affirmed and advanced modeled iterative reconstruction
iDose4, IMR, Precise ImagePhilipsHybrid, model-based and AI reconstruction
AIDR 3D, AiCECanon MedicalAdaptive iterative dose reduction and the Advanced intelligent Clear-IQ Engine deep-learning reconstruction
Intelli IP, Intelli IP QuickFujifilmThe iterative processing on the SCENARIA View and Supria families
MARMetal artifact reductionSoftware that cleans the streaks from hips, spines and dental work
Kernel, filterReconstruction algorithm sharpnessSoft kernels for soft tissue, sharp kernels for bone and lung
SNR, CNR, LCDSignal-to-noise, contrast-to-noise, low-contrast detectabilityThe image-quality measures on the phantom; what ACR accreditation checks

The brand names matter when comparing quotes because they are licensed options as well as technology. A refurbished scanner “with ASiR” is a different product from the same scanner without it, and the difference shows in dose and in image quality at low dose. Our comparison of CT scanner manufacturers places each vendor’s reconstruction in its model line, and the CT scanner cost guide explains how software licenses move a quote within its band.

GE BrightSpeed CT scanner, angled view with the table extended
A GE BrightSpeed, the 16-slice workhorse of the refurbished market: ASiR, Smart Dose features and XR-29 compliance are licensed options that may or may not be included on a given unit. Photo courtesy of GE HealthCare.
Family 4

Exam and Protocol Acronyms

TermStands forWhat it means for you
CTACT angiographyA contrast scan timed to the arteries; head, neck, chest, aorta, runoff
CTPACT pulmonary angiographyThe scan for suspected pulmonary embolism
CCTACoronary CT angiographyECG-gated cardiac CT; needs fast rotation and good temporal resolution
CACCoronary artery calcium scoreA non-contrast, low-dose axial scan scored in Agatston units
HRCTHigh-resolution CTThin-section lung imaging for interstitial disease
LDCTLow-dose CTLung cancer screening at a fraction of the dose of a diagnostic chest CT
CT KUBKidneys, ureters, bladderThe non-contrast stone scan
CTU, CTC, CTECT urography, colonography, enterographyOrgan-specific contrast protocols
With, without, with and withoutIV contrast phasesThe protocol decision explained in our guide to CT with contrast vs without
LOCM, IOCMLow-osmolar, iso-osmolar contrast mediaThe modern non-ionic iodinated agents
CT simCT simulationA wide-bore scanner with a flat table used to plan radiation therapy
PET/CT, SPECT/CTHybrid nuclear medicine scannersA CT bolted to a PET or SPECT camera; see the PET/CT scanner cost guide
Family 5

Workflow, IT, Buying and Service Terms

TermStands forWhat it means for you
DICOMDigital Imaging and Communications in MedicineThe image and communication standard; every scanner speaks it, and DICOM conformance is how you know what it can send
PACS, RIS, HIS, EHRPicture archiving, radiology, hospital and electronic health record systemsWhere the images and the orders live; a new scanner is integrated with all of them
MWL, MPPSModality worklist, modality performed procedure stepThe DICOM services that send the patient list to the scanner and report the exam back
HL7Health Level SevenThe messaging standard between the RIS, the HIS and the scanner’s worklist
OEM, ISOOriginal equipment manufacturer, independent service organizationWho built it and who services it; Medical Outfitters is an authorized Fujifilm dealer and an independent service organization for GE, Siemens, Philips and Canon
Refurbished, used, as-isCondition tiersExplained in our guide to refurbished vs OEM certified equipment; MO Select is Medical Outfitters’ refurbished line
EOL, EOS, EOGSEnd of life, end of service, end of guaranteed supportThe dates after which the manufacturer stops supporting a platform; the single most important date on a used scanner
PM, FSEPreventive maintenance, field service engineerThe scheduled visits and the people who make them; see CT scanner service
Scan seconds, tube countTube life measuresHow much of the tube’s rated life is used; explained in this guide
Tube coverageService contract termWhether the contract pays for the tube; the biggest single variable in a CT contract price
510(k)FDA premarket notificationThe clearance a scanner and its software options carry; options added later carry their own
CRCPD, state registrationConference of Radiation Control Program Directors; state radiation programsThe state side of CT regulation: registration, shielding surveys and inspections

Two of the terms in this last table decide more purchases than any of the technique or image terms. End of life, because a scanner the manufacturer no longer supports is a scanner whose parts and software are on a clock, and tube coverage, because the tube is the one bill that can equal a year of service. Both are covered in detail in our CT scanner cost and price guide, and both are stated plainly on every MO Select CT that Medical Outfitters sells.

Next Step

Talk to Medical Outfitters in Plain English

Medical Outfitters sells new Fujifilm CT scanners, including the SCENARIA View and the Supria 128, and refurbished GE and Siemens systems through MO Select, with installation, calibration, training and service by our own engineers across the United States, Puerto Rico and the Caribbean. The current systems are on the CT scanners for sale page, and our model guides for the GE LightSpeed, Optima CT660, Revolution and Canon Aquilion families translate each platform’s specification sheet into the terms above. If a quote you are reading uses an acronym this page does not, ask us; we will explain it and add it here.

Frequently Asked Questions

CT Acronyms: Common Questions

What does CT stand for?

Computed tomography: an X-ray tube and detector rotate around the patient and a computer reconstructs cross-sectional images from the measurements. CAT, computed axial tomography, is the older name for the same thing; MDCT and MSCT, multidetector and multislice CT, describe every scanner sold since the early 2000s.

What is CTDIvol, and how is it different from DLP?

CTDIvol is the dose index for one rotation of a protocol, measured in a standard phantom in milligray and shown on the console before every scan. DLP, dose length product, is CTDIvol multiplied by the scanned length, in milligray-centimeters, and represents the whole exam. Effective dose in millisieverts is estimated from DLP, and SSDE corrects CTDIvol for the patient's actual size.

What do kVp and mAs mean on a CT scanner?

kVp is the tube voltage in kilovolts, which sets how penetrating the X-ray beam is; mAs is the tube current in milliamperes multiplied by the exposure time, which sets how many X-rays are produced and is the main dose dial. Automatic exposure control varies the mA with the thickness of the patient along the scan.

What is a Hounsfield unit?

The CT number assigned to each voxel, scaled so that water is 0 and air about minus 1,000; dense bone runs from several hundred to above 1,000 and soft tissues cluster between about 20 and 80. Contrast enhancement is measured as the rise in Hounsfield units after the injection.

What are ASiR, SAFIRE, iDose and AIDR?

Vendor brand names for iterative reconstruction: ASiR, ASiR-V and TrueFidelity from GE, SAFIRE and ADMIRE from Siemens, iDose4, IMR and Precise Image from Philips, AIDR 3D and AiCE from Canon, and Intelli IP from Fujifilm. All lower image noise so the dose can be reduced, and on a refurbished scanner they are licensed options that may or may not be included.

What is XR-29, and why does it matter?

NEMA XR-29-2013, the MITA Smart Dose standard, requires four features: DICOM radiation dose structured reports, dose check alerts, automatic exposure control and reference adult and pediatric protocols. Under the Protecting Access to Medicare Act of 2014, Medicare paid 5 percent less for scans on non-compliant scanners in 2016 and 15 percent less from 2017 onward, so compliance is a line on every used CT quote.

What is the difference between slices and detector channels?

Channels are the physical detector rows the scanner reads in one rotation; slices are the images it reconstructs. On some platforms a 64-channel detector reconstructs 128 slices, so a scanner marketed by its slice count should be compared on its detector channels, its coverage and its tube as well.

What do EOL and EOS mean for a used CT scanner?

End of life and end of service, the dates after which the manufacturer stops supporting a platform with parts and software. A scanner past those dates can still be serviced by an independent organization with a parts pool, but the dates are the most important ones on a used CT quote, and Medical Outfitters states them on every MO Select system.