MRI safety zones are the four areas the American College of Radiology defines around every MRI scanner, from the public lobby to the magnet room, with screening and access control tightening at each boundary. Zone I is open to anyone. Zone II is where patients and visitors are screened. Zone III is restricted to screened people under the supervision of MR personnel, and Zone IV is the magnet room itself, where the field is always on and an unscreened steel object becomes a projectile. The zones are not a suggestion: they are the backbone of the ACR Manual on MR Safety that accreditation bodies and most state inspectors expect to see drawn on the floor plan, posted on the doors and followed by every person who walks past the reception desk.
This guide explains what each MRI safety zone is, who may enter it and under what supervision, how screening and ferromagnetic detection work at the Zone II boundary, what the 5-gauss line has to do with the zones, how the MR Safe, MR Conditional and MR Unsafe labels are used, and how the zones are designed into a new MRI suite. Medical Outfitters builds, installs and services MRI suites across the United States, Puerto Rico and the Caribbean, and the zone layout is on the drawing before the first wall goes up.
- Zone I: public space, no restrictions.
- Zone II: the screening interface; patients are questioned and supervised but not yet past the control line.
- Zone III: physically restricted; only screened people under MR personnel supervision; usually holds the control room.
- Zone IV: the magnet room, always inside Zone III, entered only under the direct supervision of Level 2 MR personnel.
- The tools: a written screening form, a verbal interview, a ferromagnetic detector at the Zone II to III boundary, a locked door, and signage that says the magnet is always on.
Why an MRI Suite Is Divided Into Zones
A superconducting MRI magnet is never off. The field that makes the images also pulls ferromagnetic objects toward the bore with a force that rises steeply in the last meter or two, and the incidents that make the news, an oxygen cylinder, a floor polisher, a wheelchair, a pair of scissors, happen when something made of steel reaches the magnet room without anyone checking it. The second hazard is less visible: a patient with an implant, a pacemaker, an aneurysm clip, a cochlear device, a fragment of metal in the eye, who enters the field without being screened. The third set of hazards lives inside the room: heating of implants and tattoos by the radiofrequency field, acoustic noise, and the cryogens that vent in a quench.
The ACR’s answer, first published in 2002 and revised through the current ACR Manual on MR Safety, is to treat the suite as a series of concentric areas with a screening step at each boundary, so that by the time a person or an object reaches the magnet room, someone trained has checked it. Our guide to MRI room shielding requirements covers the RF shield, the magnetic shielding and the 5-gauss line as construction items; this one covers the people, the doors and the procedures around them.
MRI Safety Zones I, II, III and IV Explained
Zone I: public access
Everything outside the MR environment: the building entrance, the corridors, the general waiting room. The magnetic field has no practical effect here and nobody is screened. In a well-designed suite, Zone I is where the public and unscreened staff can move freely without ever meeting a restricted door.
Zone II: the screening interface
The area between the uncontrolled public space and the strictly controlled zones beyond. Patients are greeted, questioned, given the written screening form and interviewed here; changing rooms and the patient preparation area are usually in Zone II. People in Zone II are under the supervision of MR personnel and are not free to wander toward the magnet, but they have not yet passed the control line. The reception desk that watches the door into Zone III is the practical center of this zone.
Zone III: restricted
From here on, access is physically restricted, by a locked or badge-controlled door, and limited to screened patients and to MR personnel. Unscreened people and ferromagnetic objects in Zone III can cause serious injury, because Zone III is where the fringe field becomes significant and where the door to the magnet room opens. The control room, the console and the corridor to the scanner are normally in Zone III, and MR personnel supervise everyone in it.
Zone IV: the magnet room
The scanner room itself, inside the RF-shielded enclosure, always located within Zone III and clearly marked as hazardous because of the strong magnetic field. The door carries the warning that the magnet is always on. Only screened patients accompanied by MR personnel, and MR personnel themselves, enter Zone IV, and the person responsible for the room at that moment is a Level 2 MR technologist or radiologist. During a cardiac or respiratory arrest the patient is moved out of Zone IV before resuscitation begins, because the crash cart and the code team are not screened for the magnet.

MR Personnel: Who Is Allowed Where
The ACR divides staff into three groups. Non-MR personnel are everyone who has not completed the facility’s MR safety training in the past year: most clinicians, cleaners, security, visitors, delivery drivers. They are treated as patients for screening purposes and never enter Zone III or IV unescorted. Level 1 MR personnel have completed basic MR safety training and may work in Zone III without supervision, but may not supervise others in Zone IV. Level 2 MR personnel, typically the MR technologists, the MR radiologists and the MR safety officer, have the extensive training that qualifies them to supervise everyone in Zones III and IV and to make the screening decisions. The facility’s MR Medical Director holds the responsibility for the whole program, and the ACR now recommends an MR Safety Officer and an MR Safety Expert alongside.
The practical rule a new facility has to absorb is that the zones apply to employees as much as to patients. The engineer who services the chiller, the nurse who brings a patient down, the facilities manager with a key ring: all are screened, all are escorted, and the floor polisher stays on the public side of the Zone III door.
How MRI Screening Works at the Zone II Boundary
The written form. Every patient, and every other person who will go beyond Zone II, completes a screening questionnaire covering implants, devices, surgery, metal fragments, occupational exposure to metal, pregnancy and the items in their pockets. The ACR publishes model forms, and most facilities keep theirs current with the device landscape, which changes every year.
The verbal interview. A Level 2 MR technologist goes through the form with the person, asks the follow-up questions and resolves anything uncertain before the person moves on. An implant card, a device model number or an X-ray of the orbits for a worker with a history of metal fragments are the usual next steps. Nothing is resolved by assumption.
Changing and removal. Patients change into facility clothing without metal fasteners, and everything from phones and hearing aids to hairpins and medication patches is removed or checked. Clothing with metallic fibers and some cosmetics are on the list now, which surprises people.
Ferromagnetic detection. A ferromagnetic detector, a walk-through or wall-mounted system that alarms on moving ferrous metal, sits at the Zone II to III boundary or at the Zone IV door. The ACR recommends ferromagnetic detection systems as an adjunct to screening, not a replacement for it; the detector catches the forgotten object, the form and the interview catch the implant.
Device-specific labels. Everything that enters Zone IV is classified under the ASTM F2503 system: MR Safe (a green square; no metallic, conductive or magnetic components), MR Conditional (a yellow triangle; safe within stated conditions such as a field strength and a spatial gradient), or MR Unsafe (a red circle with a bar). Implants are checked against their conditions, and equipment that lives in the magnet room, the stretcher, the IV pole, the oxygen cylinder, the monitor, is MR Safe or MR Conditional and labeled as such.

The 5-Gauss Line and Where the Zones Fall
The 5-gauss line (0.5 millitesla) is the boundary beyond which the stray magnetic field is considered safe for the general public and for devices such as older pacemakers. Regulators and the ACR expect everything inside the 5-gauss line to be a controlled area, which in practice means it must lie inside Zone IV, or at the very least inside Zone III, and never in a public corridor, a waiting room or the floor above. On a modern actively shielded 1.5T or 3T magnet the 5-gauss line sits a few meters from the isocenter and is usually contained inside the magnet room; on an older or unshielded magnet, or where the room is small, the line may extend through a wall and the facility has to make the space on the other side part of the controlled zone or add magnetic shielding. The weight and shielding of the magnets we install are covered in our guide to how much an MRI machine weighs, and the manufacturer’s site planning drawing shows the 5-gauss contour for every model.
The zones and the line are related but not the same thing. The line is a physics boundary that the manufacturer draws; the zones are an administrative boundary that the facility draws around it, with margin, doors and procedures. A suite is designed so that Zone IV contains the 5-gauss line with room to spare, Zone III wraps around Zone IV with the control room and the only door into it, and Zone II sits outside with the reception desk watching that door.
Designing the Zones Into a New MRI Suite
On Medical Outfitters’ installations the zones are drawn before the RF shield is ordered, because the door positions decide them. The practical checklist:
One door into Zone III, under observation. The reception or control desk should see it. Side doors from service corridors are the classic failure, and so is a chiller or equipment room that opens into Zone III from the outside.
The control room in Zone III, with the window on the magnet room. The technologist watches the patient and the Zone IV door from the console. An equipment room that engineers need to reach without crossing Zone IV is designed on the Zone III side.
Zone IV door: labeled, closed, interlocked where possible. The magnet-is-always-on sign, the red-and-white hazard marking, and a lock or interlock that keeps the door shut when the room is unattended. Nothing ferrous is stored inside, and the room has no ferrous fixtures.
Ferromagnetic detection at the boundary. A wall-mounted or pillar detector at the Zone II to III door or the Zone IV door, positioned so that staff see the alarm.
Zone II changing and screening space. A room to complete the form, change and store belongings, so the screening happens in one place and not in a corridor.
Emergency planning. A quench button that staff can reach, a quench vent routed outside, an evacuation route that does not pass the magnet room, and a written plan for fire, code and cryogen events agreed with the fire department before opening. Our guide to what an MRI quench is explains the cryogen side.
Signage and training. Zone signs at every boundary, the ASTM labels on equipment, the annual MR safety training for everyone who works in Zone III, and the screening forms on file. Accreditation surveys and state inspections look for all of them.

MRI Suites Built Around the Zones
Medical Outfitters designs, builds and installs MRI suites with the four zones on the drawing from the first survey: the RF-shielded magnet room, the control room with its window and single door, the screening and changing space, the ferromagnetic detector, the signage and the quench path. We install new Fujifilm MRI systems as an authorized dealer and refurbished GE and Siemens magnets through MO Select, and our MRI service team keeps the suite compliant after opening, with engineers who are screened and escorted like everyone else. The current systems are on the MRI machines page, the MRI machine cost and price guide covers what a suite build adds to the magnet, and our note on MRI installation in Puerto Rico and the Caribbean covers island projects. For a zone review of an existing suite or a layout for a new one, contact Medical Outfitters.
- MRI room shielding requirements: RF shielding, magnetic shielding and the 5-gauss line
- MRI service, repair and maintenance
- MRI machine cost and price guide (2026)
- MRI machines: new, refurbished and used
- MRI machines for sale
- What is an MRI quench?
- How much does an MRI machine weigh?
- What is an MRI cold head?
- How to move an MRI or CT scanner
- Wide-bore vs open vs closed MRI
- MRI installation in Puerto Rico and the Caribbean
- MRI, CT and X-ray machines for sale in Florida
- Request a zone review or a suite layout
MRI Safety Zones: Common Questions
What are the four MRI safety zones?
Zone I is public space with no restrictions. Zone II is the screening interface where patients and visitors are questioned and supervised. Zone III is physically restricted to screened people under the supervision of MR personnel and usually holds the control room. Zone IV is the magnet room itself, always inside Zone III, entered only under the direct supervision of Level 2 MR personnel.
Who can enter Zone IV of an MRI suite?
Screened patients accompanied by MR personnel, and MR personnel themselves, with a Level 2 technologist or radiologist responsible for the room. Non-MR personnel such as cleaners, security, clinicians and engineers are screened and escorted like patients and never enter unaccompanied; a crash team does not enter at all, because the patient is moved out of Zone IV before resuscitation.
What is the 5-gauss line in MRI?
The boundary around the magnet beyond which the stray field is below 5 gauss (0.5 millitesla), the level considered safe for the general public and for devices such as older pacemakers. Everything inside the line is a controlled area, which in practice means it must lie inside Zone IV or at least Zone III, never in a public corridor or the floor above. The manufacturer's site planning drawing shows the contour for each model.
What is the difference between MR Safe, MR Conditional and MR Unsafe?
The ASTM F2503 labels. MR Safe (a green square) means no metallic, conductive or magnetic components. MR Conditional (a yellow triangle) means safe within stated conditions, such as a maximum field strength and spatial gradient. MR Unsafe (a red circle with a bar) means the item must never enter the magnet room. Implants are checked against their conditions and everything that lives in Zone IV is labeled.
Is a ferromagnetic detector required for MRI?
The ACR recommends ferromagnetic detection systems as an adjunct to the written and verbal screening, not as a replacement, and many accrediting bodies and state programs now expect one. The detector catches the forgotten object in a pocket or on a stretcher; the questionnaire and the interview catch the implant the detector cannot see.
What is Level 1 and Level 2 MR personnel?
Level 1 MR personnel have completed basic MR safety training in the past year and may work in Zone III unsupervised, but may not supervise others in Zone IV. Level 2 MR personnel, typically MR technologists, MR radiologists and the MR safety officer, have the extensive training that qualifies them to supervise everyone in Zones III and IV and to make screening decisions. Everyone else is non-MR personnel and is screened and escorted.
Where should the control room be in an MRI suite?
In Zone III, with the window on the magnet room and a view of the Zone IV door, so the technologist supervises every entry while watching the patient. The only door into Zone III should be visible from the reception or control desk, and equipment rooms that engineers need to reach should open on the Zone III side rather than through the magnet room.
Does Medical Outfitters design the zones into a new MRI suite?
Yes. The four zones are drawn on the first survey, before the RF shield is ordered, because the door positions decide them: one observed door into Zone III, the control room with its window, the screening and changing space in Zone II, the ferromagnetic detector at the boundary, the signage, the quench path and the emergency plan. Our engineers are screened and escorted on every service visit like everyone else.